64
WATER STILLS
LABORATORY EQUIPMENT
BOECO BIDEST 4
is our Bidistiller designed to work automatically with a production of 4 liters
per hour, includes several safety systems to work continuously.
A
Produces distilled water for the most common applications in the
laboratory.
A
Boiler and Condenser are made of borosilicate glass 3.3. The heaters
are sheated with quartz glass. This is ideal for producing high quality
water, free of metal ions and pyrogens, with a conductivity 1-2 µS / cm.
A
Incorporates several security systems for automatic operation: presostat,
electrovalve and level sensor.
A
Cooling water outlet by rubber tube. The condensed water not flowing
through the cooling water outlet.
A
Energy saving through distillation of preheated cooling water.
A
Distillation process visible through the front screen.
A
The main switch and monitor drivers are on the front of the unit.
A
Control of water level automatic cut off power in case of lack of water.
A
Output of the distillate on the side of the unit.
A
Output of carbon dioxide through hole of the top of condensers
A
Suitable for wall mounting.
Safety systems
Preostat
Safety thermostat protects still in the event of water supply failure.
Electrovalve
It cuts water and electrical supply to avoid useless expenses of water
and electricity.
It returns to open water inlet if the electrical supply is enough to let the water
still to work properly.
Level device
Device to fill a container without worrying when it finished. It cuts water and
electrical supply to avoid useless expenses of water and electricity.
Specification:
Output:
approx. 4 l/h
Conductivity of distilled
water (20°C):
1-2 µS/cm
pH:
Depends on tap water inlet 5,5-7
Heater:
2 x 3.000 W
Heater element:
Quartz sheated heater
Material of glass parts:
Borosilicate glass 3.3
Water consumption:
approx. 1,2 l/min
Size:
70 x 36 x 47 cm (lxwxh)
Weight:
approx. 23 kg
Code
Description
BOE 8700410
BIDEST 4, 230V, 50/60 Hz, single phase
Note on the water output:
Depends directly on the quality of water supply and other environmental fac-
tors, maybe type III (according to ASTM). As distillates are open to the atmo-
sphere before measurements can be made, the value of conductivity is really
much less (down to 10 times) than it is really measured. Any ultra pure
water in contact with the atmosphere has a conductivity of 1-2 µS/cm and a
pH of 5, due to the small amount of CO2 (0.05 ppm) from the atmosphere.