ELECTROCHEMISTRY

BOECO PORTABLE CONDUCTIVITY/TDS

SALINITY/TEMP. METER MODEL CT-480

A

Simultaneous displays of conductivity, salinity or TDS with

temperature measurements.

A

Convenient one point calibration. Calibration data is stored in memory

and is ready for use on power up.

A

Adjustable TDS coefficient, reference temperature and temperature

coefficient

A

Uses one conductivity cell for conductivity, salinity and

TDS measurement

A

Accepts 4-wire conductivity cells with k=0,475 cell constant and 2-wire

K=0,1 cell constant for improved flexibility and accuracy

A

50 data set reviewable memory

A

Automatic shut off function (30 minutes of non-use)

A

Battery life approx. 200 -500 hours

A

Low battery indicator

A

IP65 waterproof case

Specification:

Conductivity K = 0,475

Range

Resolution

Accuracy

0,0 to 475µS/cm

0,1µS/cm ±1% of reading +2µS/cm

475 to 4750µS/cm 1µS/cm

±1% of reading +5µS/cm

4,75 to 47,50mS/cm 0,01mS/cm ±1% of reading +0,05mS/cm

47,5 to 200,0mS/cm 0,1mS/cm ±2,5% of reading +0,5mS/cm

Conductivity K = 0,1

Range

Resolution

Accuracy

0,00 to 99,99µS/cm

0,01µS/cm ±1% of reading

100 to 200µS/cm

0,1µS/cm

±1% of reading

Salinity

Range

Resolution

Accuracy

0,0 to 70,0 ppt

0,1 ppt

± 0,2 % Full Scale

Temperature

Range

Resolution

Accuracy

-10,0 to 90,0 °C

0,1 °C

±0,2 % or ±4% Full Scale

whichever is greater

Reference Temp.:

From 15,0 to 25,0°C, default is 25,0°C

Temperature Coefficient:

From 0,00 to 4,00 °C, default is 1,91°C

TDS Constant:

From 0,30 to 1,00, default is 0,65

Cell Constant

4 wire conductivity cell: K = 0,475

ATC

Thermistor, 10K ohms, at 25°C

Power

9 Volt battery

Size (LxWxD)

196x70x37 mm

Weight

0,26 kg (1 x 9V Battery included)

The CT-480 will be delivered incl. the following standard accessories:

1 x 4-wire conductivity cell 3020P with k=0,475

1 x 9 V battery, 1 x operating manual, 1 x carrying case

Code

Description

BOE 5190480 Portable Conductivity/Salinity/TDS/Temperature Meter,

Model CT-480 in carrying case with 4-wire

Conductivity Cell 3020P with 6 PIN connector,

K=0,475 (50µS - 200mS/cm)

Optional Conductivity Cell:

BOE 5190109 2-wire Conductivity Cell 109P with 6 PIN connector,

K=0,1 (0,0 - 200,0µS/cm)

LABORATORY EQUIPMENT

BOECO PORTABLE CONDUCTIVITY/

TDS/SALINITY/TEMP METER

MODEL CT-480