29
PHOTOMETER
BOECO CLINICAL PHOTOMETER
MODEL PM-51
A
Semi-automated photometric system
A
Effective temperature regulation system
A
Flexible cuvette concept, interchangeable flow-through or
standard cuvette
A
Reagent-open system with high capacity for programmable methods
A
Teach-in capability for reagent application via touchscreen
A
Minimum sipping volume 250 µl
A
Double-secured liquid control with infrared bubble detector
A
Future-proof operation, can easily be upgraded
Specification:
Type:
Semi-automatic, single-beam filter photometer
Light Source:
Halogen lamp - 12 V, 20 W with protection for lifetime
Wavelength:
340 nm - 800 nm
Wavelength Selection:
Automatic via 9-position filter wheel:
6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm, 623 nm,
3 positions for optional filter of choice
Photometric Range:
0 - 2.5 Bel
Cuvette System:
Microflow cell: 32 µl, 10 mm light path interchangeable with normal standard cuvettes
(macro or semi-micro, disposable or special optical glass) Temperature Control Internal
Peltier element, temperature variable, pre-adjusted to 25 °C, 30 °C and 37 °C
Equilibration time for aspirated reaction mixture to reach 37 °C from ambient
temperature: 15 sec
Aspiration System:
Built-in peristaltic pump driven by stepper motor programmable aspiration volume
controlled by infrared light barrier
Sipping Volume:
Minimum 250 µl, typically 500 µl up to 2000 µl
Separate setting of aspirate volume and wash volume
Operator interface:
Touchscreen, for direct functions and alphanumeric inputs
Data Presentation:
Graphic display, White characters or symbols, blue background, lighted,
resolution 240 x 126 dots.
Integrated Printer:
Thermal printer, 24 characters per line
Languages:
English, German, Spanish, French, Russian and Indonesian
Memory:
General operating software can be updated by PC
Reagent open system with capacity for up to 231 programmable methods
Import of data by touchscreen or PC
Up to 50 non-linear calibration curves with max. 20 sets of points can be stored
Data Logging:
Up to 1000 results can be saved in memory automatically
Signal Port:
1 serial port for connection to an external printer or PC (bidirectional interface)
Measurement Procedures:
A
Absorbance
A
Endpoint with standard, factor or multiple standards, with or without reagent blank and/or
sample blank
A
Bichromatic end point
A
Kinetics with standard, factor or multiple standards, with or without blank
A
Fixed time with standard, factor or multiple standards, with or without reagent blank
A
Turbidimetry with optional timer function
A
Single, double and triple determinations
A
Curve fitting for non-linear standard curves
A
Free hemoglobin in combination with optional interference filters
Measuring Time:
A
Kinetic: variable from 3 - 19 deltas, time per delta 4 - 255 sec.
A
Fixed time: variable from 0 - 65535 sec.
Delay Time:
Programmable from 0 - 65535 sec.
Mains Supply:
Range 100 VAC up to 240 VAC at 50/60 Hz
Dimensions:
Length 33 cm x Width 34 cm x Height 18 cm
Weight:
5,3 kg
Code
Description
BOE PM-51
BOECO photometer PM-51, with integrated printer, flow-through cell, standard cuvette
adapter and filter wheel (340-405-492-546-578-623 nm), 90-264 V, 50/60 Hz
LABORATORY EQUIPMENT