PHOTOMETER

BOECO CLINICAL PHOTOMETER

MODEL PM-51

A

Semi-automated photometric system

A

Effective temperature regulation system

A

Flexible cuvette concept, interchangeable flow-through or

standard cuvette

A

Reagent-open system with high capacity for programmable methods

A

Teach-in capability for reagent application via touchscreen

A

Minimum sipping volume 250 µl

A

Double-secured liquid control with infrared bubble detector

A

Future-proof operation, can easily be upgraded

Specification:

Type:

Semi-automatic, single-beam filter photometer

Light Source:

Halogen lamp - 12 V, 20 W with protection for lifetime

Wavelength:

340 nm - 800 nm

Wavelength Selection:

Automatic via 9-position filter wheel:

6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm, 623 nm,

3 positions for optional filter of choice

Photometric Range:

0 - 2.5 Bel

Cuvette System:

Microflow cell: 32 µl, 10 mm light path interchangeable with normal standard cuvettes

(macro or semi-micro, disposable or special optical glass) Temperature Control Internal

Peltier element, temperature variable, pre-adjusted to 25 °C, 30 °C and 37 °C

Equilibration time for aspirated reaction mixture to reach 37 °C from ambient

temperature: 15 sec

Aspiration System:

Built-in peristaltic pump driven by stepper motor programmable aspiration volume

controlled by infrared light barrier

Sipping Volume:

Minimum 250 µl, typically 500 µl up to 2000 µl

Separate setting of aspirate volume and wash volume

Operator interface:

Touchscreen, for direct functions and alphanumeric inputs

Data Presentation:

Graphic display, White characters or symbols, blue background, lighted,

resolution 240 x 126 dots.

Integrated Printer:

Thermal printer, 24 characters per line

Languages:

English, German, Spanish, French, Russian and Indonesian

Memory:

General operating software can be updated by PC

Reagent open system with capacity for up to 231 programmable methods

Import of data by touchscreen or PC

Up to 50 non-linear calibration curves with max. 20 sets of points can be stored

Data Logging:

Up to 1000 results can be saved in memory automatically

Signal Port:

1 serial port for connection to an external printer or PC (bidirectional interface)

Measurement Procedures:

A

Absorbance

A

Endpoint with standard, factor or multiple standards, with or without reagent blank and/or

sample blank

A

Bichromatic end point

A

Kinetics with standard, factor or multiple standards, with or without blank

A

Fixed time with standard, factor or multiple standards, with or without reagent blank

A

Turbidimetry with optional timer function

A

Single, double and triple determinations

A

Curve fitting for non-linear standard curves

A

Free hemoglobin in combination with optional interference filters

Measuring Time:

A

Kinetic: variable from 3 - 19 deltas, time per delta 4 - 255 sec.

A

Fixed time: variable from 0 - 65535 sec.

Delay Time:

Programmable from 0 - 65535 sec.

Mains Supply:

Range 100 VAC up to 240 VAC at 50/60 Hz

Dimensions:

Length 33 cm x Width 34 cm x Height 18 cm

Weight:

5,3 kg

Code

Description

BOE PM-51

BOECO photometer PM-51, with integrated printer, flow-through cell, standard cuvette

adapter and filter wheel (340-405-492-546-578-623 nm), 90-264 V, 50/60 Hz