81
CENTRIFUGES
LABORATORY EQUIPMENT
Rotors for M-240 and M-240 R
–
Cat.-No.
run-up in sec.
20
28
+4
run-down in sec.
2)
temperature °C
1)
radius in mm
85
RCF
21,382
tubes per rotor
24
_
02023 02024
with high
speeds we
recommend
to use
adapter
No. 02031
capacity in ml
1.5 0.5/0.8 0,2 /0.4 2.0
dimensions Ø x L mm
8x30/45
11x38
6x18/45 11x38
Cat.-No.
run-up in sec.
19
28
+4
run-down in sec.
2)
temperature °C
1)
radius in mm
81
RCF
20,376
tubes per rotor
30
_
n=15,000 min
-1
02430B
Cat.-No.
capacity in ml
0,5
dimensions Ø x L mm 8 x 30
Angle rotor, 30-place,
Cat. No. 02430B
n= 15,000 min
-1
max RCF 20.376
Lid 0E3243 incl.
Angle rotor, 24-place
Cat. No. 02424B
n= 15,000 min
-1
,
max RCF 21.382
Lid 0E3243 incl.
40°
45°
Cat.-No.
run-up in sec.
19
28
+4
run-down in sec.
2)
temperature °C
1)
radius in mm
57
RCF
14.338
tubes per rotor
32 4 x 8
_
n=15,000 min
-1
02418A
Cat.-No.
capacity in ml
0,2
dimensions Ø x L mm 6 x 18 PCR strips
Angle rotor, 32-place, 4x 8
for PCR tubes and strips
Cat. No. 02418A
n= 15,000 min
-1
max RCF 14,338
Optional Lid 0E3423
45°
Cat.-No.
run-up in sec.
19
28
+4
run-down in sec.
2)
temperature °C
1)
radius in mm
84
RCF
21,130
tubes per rotor
20
_
n=15,000 min
-1
02426B
Cat.-No.
capacity in ml
1,8
dimensions Ø x L mm cryo tubes
Angle rotor, 20-place
for cryo tubes
Cat. No. 02426B
n= 15,000 min
-1
max RCF 21,130
Lid 0E3243 incl.
40°
Lid (Cat, No. 02425)
with bio-containment
autoclavable
Lid (Cat. No. 02423)
with bio-containment
autoclavable,
phenol-resistant
Optional for rotor 02424B, 02430B, 02426B and 02427A
–
Cat.-No.
run-up in sec.
22
30
+4
run-down in sec.
2
temperature °C
1)
radius in mm
82
RCF
20.627
tubes per rotor
30
_
02023 02024
n=15,000 min
-1
02427A
Cat.-No.
with high
speeds we
recommend
to use
adapter
No. 02031
capacity in ml
1.5 0.5/0.8 0,2 /0.4 2.0
dimensions Ø x L mm
8x30/45
11x38
6x18/45 11x38
Angle rotor, 30-place,
Cat. No. 02427A
n= 15,000 min
-1
,
max RCF 20.627
Lid 0E3243 incl.
52,5° inner /
40° outer
n=15,000 min
-1
Cat.-No. 02424B
Angle rotor, 24-place
Cat. No. 02428
n= 15,000 min
-1
,
max RCF 21,382
Lid with bio-containment,
autoclavable, phenol-re-
sistant incl.
The special raised rim of the rotor 02428 enables
centrifugation of tubes with open lids. The rim
supports the lids so they will not rip off during
centrifugation.
Centrifugation with open tube lids is performed
for example when spinning tubes with filter
elements, so called „spin columns“
45°
Cat.-No.
run-up in sec.
20
28
+4
run-down in sec.
2)
temperature °C
1)
radius in mm
85
RCF
21,382
tubes per rotor
24
_
02023 02024
with high
speeds we
recommend
to use
adapter
No. 02031
with high
speeds we
recommend
to use
adapter
No. 02031
capacity in ml
1.5 0.5/0.8 0,2 /0.4 2.0
dimensions Ø x L mm
8x30/45
11x38
6x18/45 11x38
–
–
–
1,5
2.0
mico spin columns
11x38
n=15,000 min
-1
Cat.-No. 02428